Friday, September 22, 2017

Procedementu Analiza be'e ~ IRON(Besi)

WAP IRON DR900
1. Start program 265 Iron FerroVer
2. Prepare 2,5 ml of water sample
3. add treated water to sample up to 25 ml
4. Stir a minute to make sure that it’s really mixing
5. Fill a sample cell with 10 ml of sample
6. Add the contents of one FerroVer Iron Reagent Powder Pillow.
7. Close the sample cell. Shake to dissolve the reagent.
8. Start the instrument timer. To prevent low results leave the sample on a
flat surface.
9. Fill a blank second sample cell with 10 ml of sample.
10.Clean the blank.
11.Insert the blank into the cell holder.
12.Push Zero.
13.After the timer expires, gently invert the prepared sample two times.
14.Clean the prepared sample.
15.Insert the prepared sample into the cell holder.
16.Push read.
17.Result show in Fe- times by 10
Note: 10 values are ten times mixing between water sample and
treated water.

Konteudu seluk tan bele hare' iha link ida ne': pajina Kualidade be' Betano Dadulas